15. Radiometer settings ABL800 FLEX Operator's Manual
15-6
Radiometer default settings, Continued
Temperature Corrected Values
pH(T)
pCO
2
(T)
pO
2
(T)
< New Group >
Oxygen Status
ctO
2
p50
< New Line >
Acid Base Status
cBase(Ecf)
)stP,(
3
HCOc
Default settings for Drift tolerances are as follows:
Parameter Drift 1 Drift 2 Units
pH 0.020 0.020
pCO
2
2.5 5 mmHg
pO
2
6 6 mmHg
ctHb 10 N/A pA
cK
+
0.2 1.5 mmol/L
cNa
+
3 1 mmol/L
cCa
2+
0.05 0.2 mmol/L
cCl
-
2 3 mmol/L
cGlu 0.5 N/A mmol/L
cLac 0.2 N/A mmol/L
cCrea 15 15
mol/L
Continued on next page
Analysis Setup
defaults
(continued)
Calibration
Setup defaults