15. Radiometer settings ABL800 FLEX Operator's Manual
15-12
Radiometer default settings, Continued
Default settings for Corrective Actions are as follows:
Event Default setting Traffic light
Calibration error(s) present “? on specific parameters”
Yellow
Calibration schedule reminder(s) No setting
Yellow
QC error(s) present No setting
Green
QC schedule reminders No setting
Yellow
Replacement schedule reminders No setting
Yellow
System message(s) present No setting
Yellow
User Activity Reminder(s) No setting
Yellow
AutoCheck QC error(s) present No setting
N/A
AutoCheck scheduled level missing No setting
Green
Default settings for Miscellaneous Setup are as follows:
Event Default setting
Analyzer locked Not set
Use safeTIPCAP Yes
Enable Estimated Derived Parameters No
Fixed pO
2
/pCO
2
decimals No
Enable General Barcode Support Yes
Enable Patient Result Approval No
Apply Parameter Corrections to QC Yes
Log All Measurement Activities No
Show Parameter Bar Yes
Auto temp unit conversion No
HbF correction "Enabled for levels > 20 %"
Analyzer Message (Blank)
Screen saver Yes, Minutes to wait when idle: 5
Default settings for Automatic Printing are as follows:
Item Default setting
Patient results On
QC results Off
Calibration results Off
Activity log messages Off
FLEXQ registration receipt Off
Message level User
Number of copies 1