EasyManuals Logo

RKI Instruments GX-6000 User Manual

RKI Instruments GX-6000
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
GX-6000 Operator’s Manual Replacing a Sensor • 122
fitting connections need to be facing up and cannot be skewed sideways.
The gasket must be pushed down in the center and secured under the
gasket aligning tabs.
Failure to install the sensor gasket appropriately may result in
inaccurate gas readings.
Figure 26: Sensor Gasket Seating
9 . Insert the flow chamber back into the instrument.
10 . Tighten the flow chamber’s four screws that were loosened in Step 2.
11 . Start up the GX-6000 by pressing and briefly holding the POWER/
ENTER button.
12 . To verify that the sensor gasket was inserted properly:
a. Plug the inlet with your finger.
b. Verify that the GX-6000 goes into low flow alarm. If the GX-6000
does not go into low flow alarm, turn the instrument off and attempt
to seat the sensor gasket again.
Gasket Aligning Tabs
Flow Fitting Connections

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals