GX-6000 Operator’s Manual Using Maintenance Mode 156
Updating the Bump Check Gas Setting
This item defines which sensors are used to update the bump test reminder
screen. With BUMP CHECK GAS set to 4 GAS (factory setting), the bump
test dates for only the standard 4 sensors (LEL, O2, H2S, and CO) are used to
determine if a bump test is due.
With BUMP CHECK GAS set to ALL GAS, the bump test dates for the
standard 4 sensors along with the PID sensor and any other installed smart
sensors are used to determine if a bump test is due.
1 . From the BUMP SETTING screen, place the cursor next to BUMP
CHECK GAS.
2 . Press and release POWER/ENTER. The current Bump Check Gas
setting will be displayed.
3 . Use AIR or SHIFT (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument
will return to the BUMP SETTING screen.
Returning to the Maintenance Mode Menu
1 . From the BUMP SETTING screen, place the cursor next to ESCAPE.
2 . Press and release POWER/ENTER. The instrument will return to the
main menu.
Updating Alarm Parameters
The ALARM SETTING menu item in Maintenance Mode has a sub menu
with 4 menu items: ALARM POINTS, ALARM LATCHING, ALARM
SILENCE, and ESCAPE.
1 . From the main menu, place the cursor next to ALARM SETTING.
2 . Press and release POWER/ENTER. The ALARM SETTING menu
appears.
MAINTE
ALARM SETTING
ALARM POINTS
ALARM LATCHING
ALARM SILENCE
ESCAPE