GX-6000 Operator’s Manual Tips for Using Maintenance Mode • 144
Tips for Using Maintenance Mode
• When in the main menu, the cursor (>) indicates which menu item will be
selected if the POWER/ENTER button is pressed and released.
• Use the SHIFT ▼ (PANIC) button to move the cursor down through the
main menu and submenu items, and to lower values or change the setting
in a specific option.
• Use the ▲AIR button to move the cursor up through the main menu and
submenu items, and to raise values or change the setting in a specific
option.
• Use the POWER/ENTER button to enter a menu item with the cursor next
to it and to enter and save settings during programming.
• An adjustable parameter that is flashing can be adjusted with the ▲AIR
and SHIFT ▼ (PANIC) buttons.
• Press the DISP/LOCK button while in a screen where you are entering or
updating parameters to exit the screen without saving any changes. You
can also use the DISP/LOCK button to back out of submenus and return
to the main menu.
Using Maintenance Mode
WARNING: The GX-6000 is not in operation as a gas detector while in
Maintenance Mode.
1 . Take the GX-6000 to a non-hazardous location and turn it off if it is on.
2 . Press and hold the ▲ AIR and SHIFT ▼ (PANIC) buttons, then press and
hold the POWER/ENTER button. When you hear a beep, release the
buttons.
START MEASURE (pg.174) enter Measuring Mode N/A
*
PID channel’s zero suppression is always 0% regardless of what value is shown.
Table 17: Maintenance Mode Menu Items
Menu Item (Page # of Description) Choices or Action Factory Setting