151 • Using Maintenance Mode GX-6000 Operator’s Manual
Updating the Calibration Check Gas Setting
This item defines which sensors are used to update the calibration reminder
screen. With CAL CHECK GAS set to 4 GAS (factory setting), the
calibration dates for only the standard 4 sensors (LEL, O2, H2S, and CO) are
used to determine if a calibration is due.
With CAL CHECK GAS set to ALL GAS, the calibration dates for the
standard 4 sensors along with the PID sensor and any other installed smart
sensors are used to determine if a calibration is due.
1 . From the CAL SETTING screen, place the cursor next to CAL
CHECK GAS.
2 . Press and release POWER/ENTER. The Calibration Reminder Screen
appears.
3 . Use â–²AIR or SHIFT â–¼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
CAL SETTING menu.
Returning to the Maintenance Mode Menu
1 . From the CAL SETTING screen, place the cursor next to ESCAPE.
2 . Press and release POWER/ENTER. The instrument will return to the
main menu.
Updating Bump Test Settings
The CAL SETTING menu item in Maintenance Mode has a sub menu with
7 menu items: BUMP DISP, BUMP PARAMETERS, BUMP
REMINDER, BUMP INTERVAL, BUMP EXPRD, BUMP CHECK
GAS, and ESCAPE.
1 . From the main menu, place the cursor next to BUMP SETTING.
2 . Press and release POWER/ENTER. The BUMP SETTING menu
appears.
MAINTE
BUMP SETTING
BUMP DISP
BUMP PARAMETERS
BUMP REMINDER
BUMP INTERVAL
BUMP EXPRD
BUMP CHECK GAS
ESCAPE