97 • Entering User Mode GX-6000 Operator’s Manual
Entering User Mode
WARNING: The GX-6000 is not in operation as a gas detector while in
User Mode.
1 . Take the GX-6000 to a non-hazardous location and turn it off if it is on.
2 . Press and hold the â–² AIR and SHIFT â–¼ (PANIC) buttons, then press and
hold the POWER/ENTER button. When you hear a beep, release the
buttons.
3 . The screen that appears will depend on the setting of User Mode’s
PASSWORD item.
If PASSWORD is set to OFF (factory setting), continue with Step 4.
If PASSWORD is set to ON, continue with Step 6.
4 . If PASSWORD has been set to OFF, the Mode Select screen will
appear, prompting you to choose the mode you’d like to enter.
5 . Use the â–²AIR and SHIFT â–¼ (PANIC) buttons to put the cursor in front
of USER MODE and press and release the POWER/ENTER button.
Continue to Step 8.
6 . If PASSWORD has been set to ON in User Mode, a password screen
will appear. The first digit will be flashing.
7 . Enter the password by using the â–²AIR and SHIFT â–¼ (PANIC) buttons
to select each password number and then pressing and releasing the
POWER/ENTER button to enter it and move on to the next number until
all of the numbers are entered. Continue to Step 8.
PASSWORD
MODE SELECT
USER MODE
MAINTENANCE
PASSWORD
INPUT
PASSWORD
0 0 0 0
PASSWORD set to OFF PASSWORD set to ON

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals