GX-6000 Operator’s Manual Display Mode 54
Pump Off Screen
WARNING: The GX-6000 is not a gas monitoring device while the pump
is off.
NOTE: The item appears in Display Mode only if the PUMP OFF DISP
menu item in Maintenance Mode is turned ON. The factory setting
for PUMP OFF DISP is OFF.
This feature can be used to conserve battery life if there is a significant period
of time until the next gas monitoring task, but you do not want to wait for the
warm-up sequence to complete if you turn the instrument off and on again.
1 . In this screen, press and release the POWER/ENTER button to turn the
pump off. The fan symbol in the upper right corner will disappear. Once
the pump is off, you cannot exit this screen and return to Measuring
Mode or continue in Display Mode until you turn the pump back on.
2 . To turn the pump back on and continue in Display Mode or return to
Measuring Mode, press and release the DISP/LOCK button.
DISP
PUMP OFF
YES :ENTER
NO :DISP
EXIT :RESET
1:44
PUMP OFF
OFF