GX-6000 Operator’s Manual Calibrating Using the Auto Calibration Method • 74
Performing a Span Adjustment
1 . Install the demand flow regulator onto the 4-gas or 5-gas calibration
cylinder.
2 . Connect the sample tubing to the demand flow regulator.
3 . Install the sample hose and probe on the GX-6000 inlet fitting. Make
sure the probe’s two halves are tightened firmly together to avoid leaks
that can affect the calibration. See Figure 22, “Replacing the Particle
Filter and Hydrophobic Filter Disk” on page 116 for an illustration of the
internal parts of the probe.
4 . Move the cursor next to the AUTO CAL menu item by using the AIR
or SHIFT (PANIC) button.
5 . Press and release the POWER/ENTER button to display the Auto Cal
Menu.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE