123 • Replacing the Hydrophobic Filter and Wire Mesh Disk GX-6000 Operator’s
13 . Calibrate the new sensors as described inChapter 4: Calibration Mode”
on page 68.
Replacing the Hydrophobic Filter and
Wire Mesh Disk
1. Verify that the GX-6000 is off.
2. Locate the clear plastic filter holder at the top of the GX-6000.
3. Grasp the filter holder and turn it 1/4 turn counterclockwise.
4. Pull the filter holder away from the case.
5. The wire mesh disk and hydrophobic filter are located in the case and are
retained by a rubber gasket.
Figure 27: Changing the Filter & Wire Mesh Disk
Rubber Gasket
(Inside Instrument)
Filter Holder
Hydrophobic
Filter
Wire Mesh Disk