EasyManua.ls Logo

RKI Instruments GX-6000 - Calibrating the ESS-03 Sensor Using Single Calibration

RKI Instruments GX-6000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
189 ESS-03 Calibration GX-6000 Operator’s Manual
If the ESS-03 channel cannot be adjusted to the proper value, a screen
displays that indicates a calibration failure and indicates that the ESS-03
sensor failed to calibrate.
The buzzer and alarm LEDs activate in a double pulsing pattern. Press
and release the RESET button to reset the alarm and continue to the Auto
Cal Menu. Attempt to calibrate the ESS-03 channel again. If the failure
continues, investigate the cause. See “Troubleshooting” on page 104.
16 . Disconnect the tubing from the probe.
17 . Unscrew the demand flow regulator from the calibration cylinder.
18 . Use the SHIFT (PANIC) button to move the cursor next to ESCAPE,
then press and release the POWER/ENTER button to return to the
Calibration Mode Menu.
19 . Use the SHIFT (PANIC) button to place the cursor next to the
NORMAL MODE menu option, then press and release the POWER/
ENTER button to return to Measuring Mode.
Calibrating the ESS-03 Sensor Using Single Calibration
See “Performing a Span Adjustment” on page 74 for instructions to calibrate
the ESS-03 using single calibration.
NOTE: RKI Instruments, Inc. recommends that you dedicate a regulator for
use with chlorine (Cl
2
) gas and that you do not use that dedicated
regulator for any other gases, particularly hydrogen sulfide (H
2
S).
APPLYGAS
HCN
ppm
FAIL

Table of Contents

Other manuals for RKI Instruments GX-6000

Related product manuals