GX-6000 Operator’s Manual Start Up and Operation 224
NOTE: If INERT MODE is set to OFF in Maintenance Mode, the INERT
item will not appear in the Alarm Points Menu.
2 . Select INERT (not O2) in order to view or change Inert Mode’s oxygen
alarm points.
Start Up and Operation
1 . Connect the tapered rubber nozzle or the sample hose and probe to the
GX-6000’s quick connect inlet fitting.
2 . Press and briefly hold down the POWER/ENTER button. Release the
button when you hear a beep.
3 . The Normal Mode Select Screen displays. Use the AIR or SHIFT
(PANIC) button to scroll to the Inert Mode Select Screen.
NOTE: INERT MODE must be set to ON in Maintenance Mode for Inert
Mode to be an option in the Mode Select Screen at startup.
4 . With the Inert Mode Select Screen displayed, press and release the
POWER/ENTER button to begin the Inert Mode startup sequence.
ALARM - P
ALARM POINTS
CH4
O2
H2S
CO
VOC
INERT
ESCAPE
4:07
MODE SELECT
NORMAL MODE
CH4 100 %LEL
O2 40.0 %
H2S 100.0 ppm
CO 500 ppm
VOC 6000 ppm
4:07
MODE SELECT
INERT MODE
AIR or
SHIFT (PANIC)