45 • Display Mode GX-6000 Operator’s Manual
PID Gas Name Screen
The standard PID channel is configured for and calibrated to isobutylene. If
calibration to a different gas is required for an application, the PID channel
can also be configured for other gases. Regardless of what gas the PID sensor
is configured for and calibrated to, the sensor will still detect and respond to a
variety of volatile organic compounds (VOCs).
If you configure the instrument for another gas, the instrument will retain that
configuration until you change it again. Turning the instrument off and on
does not change this setting.
1 . With the PID Gas Name Screen displayed, press and release the
POWER/ENTER button. The top level of the PID Gas Name List will be
displayed and the cursor will be next to the letter corresponding to the
first letter of the current setting. In the example below, the PID channel
is configured for isobutylene so the cursor is next to “I”. Each letter has a
list of gas names behind it. At the top of the letter list, there is a User List
item and a Recent List item. The User List can hold 30 gas names and
the GX-6000 Data Logger Management Program is required to
configure this list. The Recent List can hold up to 8 gas names and is
made up of the 8 most recently used gas names. The Recent List is not
user editable.
DISP
SELECT PID LIST
PID SET TO
Isobutylene
10.6 eV
CHANGE :ENTER
KEEP :DISP
EXIT :RESET
DISP
SELECT PID 11/24
F -
G -
H -
I-
J -
K -
L -
M -