161 • Using Maintenance Mode GX-6000 Operator’s Manual
3 . Use â–²AIR or SHIFT â–¼ (PANIC) to move the cursor in front of the
desired language.
If you do not wish to select a new language, press and release DISP/
LOCK. The unit will return to the main menu.
4 . Press and release POWER/ENTER to save the new language setting and
return to the main menu. The GX-6000’s user interface will now be in
the newly selected language.
5 . If you select a language other than English, a prompt will appear during
startup that allows you to change the language back to English if desired.
Updating the Lunch Break Setting
With LUNCH BREAK set to OFF (factory setting), the GX-6000
automatically starts new TWA and PEAK reading collection and resets the
time in operation at startup.
With LUNCH BREAK set to ON, the Lunch Break Screen displays during
startup. From this screen, you can choose to continue accumulating TWA and
PEAK readings and the time in operation from the last time the GX-6000 was
used or start collecting new readings and reset the time in operation.
1 . From the main menu, place the cursor next to LUNCH BREAK.
2 . Press and release POWER/ENTER. The Lunch Break Screen appears.
3 . Use â–²AIR or SHIFT â–¼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
main menu.
Updating the Auto Zero Setting
This setting allows you to configure the GX-6000 so that a fresh air
adjustment takes place automatically as part of the instrument startup
sequence. If AUTO ZERO is set to ON, the GX-6000 performs a fresh air
adjustment at the end of the startup sequence before entering Measuring
Mode. The factory setting is OFF.
WARNING: If the automatic fresh air feature is turned on, you must
startup the GX-6000 in a known fresh air environment, an
environment free of toxic or combustible gases and of normal
oxygen content (20.9%). If this feature is on and the GX-6000
is started up in the presence of a target gas, the readings and
alarms will not be accurate or reliable.
1 . From the main menu, place the cursor next to AUTO ZERO.

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals