47 • Display Mode GX-6000 Operator’s Manual
7 . To return to the PID Gas Name Screen, press and release the DISP/
LOCK button one more time.
Peak Screen
The peak screen displays the highest (lowest for oxygen) concentrations
detected since the GX-6000 was turned on. Peak readings are stored in the
GX-6000’s memory until a higher level is detected (lower for oxygen), the
peak reading is cleared, or the GX-6000 is turned off.
The lunch break feature enables the GX-6000 to save peak readings when it
is turned off so it can continue them when it is turned on again. See “Turning
On the GX-6000” on page 19.
To clear the peak readings, do the following:
1 . With the Peak Screen displayed, press and hold the RESET button. The
following screen sequence will occur.
2 . Release the RESET button. The peak readings will be reset and the unit
will return to the Peak Screen.
If you do not want to clear the peak readings, release the RESET button
before the above screen sequence occurs. The unit will return to the Peak
Screen.
PEAK
CH4 O2
%LEL %
H2S CO
ppm ppm
VOC
ppm
10 15.0
13.0 200
2000
PEAK
PEAK CLEAR
HOLD
PEAK
PEAK CLEAR
RELEASE