GX-6000 Operator’s Manual Overview • 140
Appendix A: Maintenance Mode
Overview
This appendix describes the GX-6000 in Maintenance Mode. The GX-6000
is factory-set to suit most applications. Update settings in Maintenance Mode
only if required for your specific application. Maintenance Mode items and
their factory settings are listed in Table 17 below.
Table 17: Maintenance Mode Menu Items
Menu Item (Page # of Description) Choices or Action Factory Setting
DATE (pg.147) user-entered date/time current date/
time
DATE FORMAT (pg.148) DD/MM/YYYY
MM/DD/YYYY
YYYY/MM/DD
MM/DD/YYYY
GAS CALIBRATION (pg.148) perform a calibration N/A
BUMP TEST (pg.148) perform a bump test N/A
CAL SETTING (pg.149)
• CAL REMINDER (pg.149) ON
•OFF
ON
• CAL INTERVAL (pg.149) 1-365 DAYS 90 DAYS
• CAL EXPRD (pg.150) CONFIRM TO USE
CANNOT USE
•NO EFFECT
CONFIRM TO
USE
• CAL CHECK GAS (pg.151) 4 GAS
•ALL GAS
4 GAS
• ESCAPE return to main menu N/A
BUMP SETTING (pg.151)
• BUMP DISP (pg.152) ON
•OFF
OFF
• BUMP PARAMETERS (pg.152)
• GAS TIME (pg.152) 30 SEC
45 SEC
60 SEC
90 SEC
30 SEC