GX-6000 Operator’s Manual Using Maintenance Mode • 170
Returning to the Maintenance Mode Menu
1 . From the LOG SETTING screen, place the cursor next to ESCAPE.
2 . Press and release POWER/ENTER. The instrument will return to the
main menu.
Adjusting the Low Flow Setpoint
If your instrument repeatedly goes into low flow alarm and you have tried all
of the recommendations in the Troubleshooting section, you may need to
perform a low flow setpoint adjustment.
1 . Use the ▲AIR or SHIFT ▼ (PANIC) button to place the cursor next to
FLOW ADJUST.
2 . Press and release the POWER/ENTER button. The pump will turn off
and the screen will display values that reflect the pump’s current draw.
3 . Use the ▲AIR and SHIFT ▼ (PANIC) buttons to get a reading of
approximately 1500 on the bottom value. When you have adjusted the
value as close as you can, press and release the POWER/ENTER button.
4 . The pump will turn on and two new values will be displayed. The top
will be a reference value and the bottom will reflect the pump’s current
draw.
MAINTE
FLOW ADJUST
50
1500
MAINTE
FLOW ADJUST
1500
2500