169 • Using Maintenance Mode GX-6000 Operator’s Manual
Updating the Data Log Interval Setting
This setting indicates how often the GX-6000 saves readings to the data
logger. The following interval times can be selected: 10 minutes, 5 minutes
(factory setting), 3 minutes, 2 minutes, 1 minute, 30 seconds, 20 seconds, or
10 seconds.
1 . From the LOG SETTING screen, place the cursor next to LOG
INTERVAL.
2 . Press and release POWER/ENTER. The Log Interval Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
LOG SETTING screen.
Updating the Log Clear Display Setting
With LOG CLEAR DISP set to ON (factory setting), the LOG CLEAR
Screen appears in Display Mode (see “LOG CLEAR Screen” on page 52).
With LOG CLEAR DISP set to OFF, the LOG CLEAR Screen does not
appear in Display Mode.
1 . From the LOG SETTING screen, place the cursor next to LOG
CLEAR DISP.
2 . Press and release POWER/ENTER. The Log Clear Display Screen
appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
LOG SETTING screen.
Updating the Data Log Overwrite Setting
With LOG OVER WRITE set to ON (factory setting), the GX-6000 writes
over the oldest data with new data when the data logger memory is full.
With LOG OVER WRITE set to OFF, the GX-6000 stops saving data to the
data logger when the data logger memory is full.
1 . From the LOG SETTING screen, place the cursor next to LOG OVER
WRITE.
2 . Press and release POWER/ENTER. The Log Over Write Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
LOG SETTING screen.