GX-6000 Operator’s Manual Measuring Mode, Normal Operation • 34
Snap Log Mode
The snap logging function in Snap Log Mode allows the user to record data at
a specific time and have it saved to the data logger. The data is assigned a
snap log ID and is saved with the station ID that was in use when the data
was taken.
To enter Snap Log Mode and record snap log data:
1 . While the unit is in Measuring Mode, press and hold the SHIFT â–¼
(PANIC) button, then press and hold the â–²AIR button and hold both
until you hear a beep. The unit will cycle through the following scre e ns.
The first screen displays what snap log ID will be given to this particular
set of data and that you should press the POWER ENTER button to save
a set of snap log data. The snap log ID number increases sequentially
with each set of snap log data taken. The second screen displays what
Station ID will be associated with this snap log and that you should press
the DISP/LOCK button to exit Snap Log Mode without saving a set of
snap log data. The third screen displays the year, month, day, and time of
the snap log. The fourth screen displays the current gas readings.
REC DATA
REC. DATA
NUMBER 001
YES : ENTER
REC DATA
REC. DATA
STATION ID
TANK 5
NO : DISP
REC DATA
REC. DATA
05/20/2015
MM/DD/YYYY
3:38
REC DATA
CH4 O2
%LEL %
H2S CO
ppm ppm
VOC
ppm
0 20.9
0.0 0
0

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals