EasyManua.ls Logo

RKI Instruments GX-6000 - Replacing an IR Sensor

RKI Instruments GX-6000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GX-6000 Operator’s Manual Replacing an IR Sensor 200
Replacing an IR Sensor
When replacing a sensor, you may either replace it with another of the same
sensor or you may install a different IR sensor. If a different one is installed,
the GX-6000 will load the sensor parameters and configure the IR channel
for the new sensor.
1 . Follow the instructions listed in “Replacing a Sensor” on page 120.
Be sure that the new sensor is installed in the same position as the old sensor
and that the following sensor priority for the smart sensor 1 position is
maintained.
Table 26: Priority for Smart Sensor 1 Position
For example, if your instrument is a 4-gas + high range PID + IR HC, the
high range PID sensor should be in the smart sensor 1 position and the IR HC
sensor should be in the smart sensor 2 position. If your instrument is a 4-gas
+ IR HC, the IR HC sensor should be in the smart sensor 1 position and a
dummy sensor should be installed in the smart sensor 2 position. See the
figure below for the location of the smart sensor 1 and 2 positions.
High Priority
Low Priority
Low range PID
10.0 eV/benzene PID
High range PID
ESS-03 Chlorine
ESS-03 Ammonia
IR and other ESS-03 sensors

Table of Contents

Other manuals for RKI Instruments GX-6000

Related product manuals