GX-6000 Operator’s Manual Using Maintenance Mode 2 178
10 . The main menu displays.
11 . Use the AIR or SHIFT (PANIC) button to move the cursor up and
down through the menu items.
Changing the Gas Combination
The GAS COMB menu item allows you to turn channels on and off.
NOTE: A smart sensor channel cannot be turned off if the smart sensor is
still installed. If you want to turn off a smart sensor’s channel, you
must first remove the smart sensor and replace it with a dummy
sensor.
1 . From the main menu, place the cursor next to GAS COMB.
2 . Press and release POWER/ENTER. The Gas Combination Screen
appears and all channels are displayed.
3 . Move the cursor next to the channel you want to update. Press and
release POWER/ENTER. The channel’s current on/off setting will begin
to flash.
4 . Use AIR and SHIFT (PANIC) to change the on/off setting.
MAINTE 2
MAINTENANCE 2
GAS COMB
HC SELECT
START MEASURE
FLASH
GAS COMB
HC ON
OX ON
H2S ON
CO ON
SM1 ON
SM2 OFF
ESCAPE