GX-6000 Operator’s Manual Using Maintenance Mode • 148
Setting the Date Format
The date can be displayed in three ways, MM/DD/YYYY (factory setting),
DD/MM/YYYY, or YYYY/MM/DD.
1 . From the main menu, place the cursor next to DATE FORMAT.
2 . Press and release POWER/ENTER. The Date Format screen appears
with the current setting displayed.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
main menu.
Performing a Calibration
1 . From the main menu, place the cursor next to GAS CALIBRATION.
2 . Press and release POWER/ENTER. The Gas Calibration Screen appears.
3 . Use the ▲AIR or SHIFT ▼ (PANIC) buttons to move the cursor to the
operation you wish to perform.
See “Calibrating Using the Auto Calibration Method” on page 71 for
instructions to perform an automatic calibration.
See “Calibrating Using the Single Calibration Method” on page 84 for
instructions to perform a single calibration.
Performing a Bump Test
NOTE: The BUMP DISP menu item in Maintenance Mode is factory set to
OFF. The BUMP TEST menu item will not appear unless BUMP
DISP is set to ON. See “Updating the Bump Display Setting” on
page 152 for instructions.
Entering the BUMP TEST menu item brings you to the first of the Gas
Select Screens. See “Performing a Bump Test” on page 90 for instructions,
starting at Step 8.
MAINTE
GAS CALIBRATION
AIR CAL
AUTO CAL
SINGLE CAL
ESCAPE