149 • Using Maintenance Mode GX-6000 Operator’s Manual
Updating Calibration Settings
The CAL SETTING menu item in Maintenance Mode has a sub menu with
5 menu items: CAL REMINDER, CAL INTERVAL, CAL EXPRD, CAL
CHECK GAS, and ESCAPE.
1 . From the main menu, place the cursor next to CAL SETTING.
2 . Press and release POWER/ENTER. The CAL SETTING menu appears.
Updating the Calibration Reminder Setting
With CAL REMINDER set to ON (factory setting), the GX-6000 will give
an indication at start up if it is due for calibration. The type of indication will
depend on the CAL EXPRD setting (see “Updating the Calibration Expired
Setting” on page 150).
With CAL REMINDER set to OFF, the GX-6000 will not give an
indication at start up if it is due for calibration.
1 . From the CAL SETTING screen, place the cursor next to CAL
REMINDER.
2 . Press and release POWER/ENTER. The Calibration Reminder Screen
appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
CAL SETTING menu.
Updating the Calibration Interval
This setting defines the amount of time between calibrations. The time can be
set in 1 day increments. The minimum setting is 1 day and the maximum
setting is 365 days. The factory setting is 90 days.
1 . From the CAL SETTING screen, place the cursor next to CAL
INTERVAL.
MAINTE
CAL SETTING
CAL REMINDER
CAL INTERVAL
CAL EXPRD
CAL CHECK GAS
ESCAPE