GX-6000 Operator’s Manual Using Maintenance Mode 2 • 176
Using Maintenance Mode 2
WARNING: The GX-6000 is not in operation as a gas detector while in
Maintenance Mode 2.
1 . Take the GX-6000 to a non-hazardous location and turn it off if it is on.
2 . Press and hold the â–² AIR and SHIFT â–¼ (PANIC) buttons, then press and
hold the POWER/ENTER button. When you hear a beep, release the
buttons.
3 . The screen that appears will depend on the setting of User Mode’s
PASSWORD item.
If User Mode’s
PASSWORD is set to OFF (factory setting), continue
with Step 4.
If User Mode’s
PASSWORD is set to ON, continue with Step 6.
4 . If User Mode’s
PASSWORD has been set to OFF, the Mode Select
screen will appear, prompting you to choose the mode you’d like to
enter.
5 . Use the â–²AIR and SHIFT â–¼ (PANIC) buttons to put the cursor in front
of MAINTENANCE and press and release the POWER/ENTER button.
A password screen will appear. Continue to Step 7.
6 . If User Mode’s
PASSWORD has been set to ON, a password screen will
appear.
PASSWORD
MODE SELECT
USER MODE
MAINTENANCE
PASSWORD
INPUT
PASSWORD
0 0 0 0
User PASSWORD OFF User PASSWORD ON User PASSWORD ON
Maint
. PASSWORD ON Maint. PASSWORD ON Maint. PASSWORD OFF
or OFF
4:07
INPUT
PASSWORD
- - - -
ENTER FOR
MAINTENANCE MODE