EasyManua.ls Logo

RKI Instruments GX-6000 - Gas Display Screen

RKI Instruments GX-6000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61 Display Mode GX-6000 Operator’s Manual
3 . Press and release the POWER/ENTER button to save the setting and
return to the Peak Bar Screen.
To return to the Peak Bar Screen without saving any changes, press and
release the DISP/LOCK button.
Gas Display Screen
The Gas Display Screen allows the user to choose how the gas readings are
displayed in Measuring Mode. A description of the choices can be found in
Table 10 below. If anything other than DISPLAY ALL is selected, the
instrument will return to DISPLAY ALL when it is turned off and turned
back on.
Table 10: Gas Display Choices
Option Description
DISPLAY ALL All active channels’ readings will be displayed at once
SCROLL AUTO One active channel will display at a time and the instrument will
automatically scroll through all active channels for the duration of
the operating session.
SCROLL MANUAL One active channel will display at a time and the user must press
and release the POWER/ENTER button to scroll through all active
channels. If the POWER/ENTER button is not pressed and
released, the instrument will display the same channel’s readings
for the duration of the operating session.
DISP
GAS DISP
YES :ENTER
NO :DISP
EXIT :RESET

Table of Contents

Other manuals for RKI Instruments GX-6000

Related product manuals