GX-6000 Operator’s Manual Calibrating Using the Single Calibration Method 84
Calibrating Using the Single
Calibration Method
Single Calibration allows you to calibrate one channel at a time. This is
useful if you only want to calibrate one or two channels.
Setting the Fresh Air Reading
1 . While in the Calibration Mode Screen, move the cursor to the AIR CAL
menu item by using the SHIFT (PANIC) button.
2 . Press and release the POWER/ENTER button. The Fresh Air Reading
Screen will be displayed.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
AIR
CH4 O2
%LEL %
H2S CO
ppm ppm
VOC
ppm
0 20.9
0.0 0
0