GX-6000 Operator’s Manual Leak Testing 214
Snap log data that was taken in Leak Check Mode operation will have 2
sequential screens that are part of one data set. One screen is for the base
reading and the other screen is for the peak reading. “BASE” and
“PEAK” will appear in the gas reading screen for each type of reading,
respectively. If a base data point appears without a peak data point after
it, it is because a peak data point was not taken.
NOTE: Even if CO DISPLAY was set to ON when the Leak Check snap
logs were taken, the CO reading will not be displayed. The CO
readings can be viewed using the GX-6000 Datalogging Program.
7 . Use the AIR and SHIFT (PANIC) buttons to scroll through the
different snap log data screens. The gas readings will change but the
snap log ID is not visible from this screen.
8 . You can also go back and forth between the ID and data screens by
pressing and releasing the POWER/ENTER button.
9 . To return to Leak Check Mode, press and release the DISP/LOCK
button.
Turning Off the GX-6000, Leak Check Mode
1 . Press and hold the POWER/ENTER button.
2 . TURN OFF will appear on the display and the buzzer will pulse for
about five seconds.
3 . Release the button when TURN OFF disappears from the display.