GX-6000 Operator’s Manual Using Maintenance Mode • 168
4 . Press and release POWER/ENTER to save the setting and return to the
MAN DOWN MENU screen.
Updating the ALARM TIME Setting
The ALARM TIME is the amount of time that has to pass between a Man
Down detection and the main alarm. It can be set in 1 second increments
from 10 - 120 seconds. The factory setting is 90 seconds. When setting the
ALARM TIME, keep in mind that ALARM TIME WARNING 2
TIME WAR NIN G 1 TI ME .
1 . From the MAN DOWN MENU screen, place the cursor next to
ALARM TIME.
2 . Press and release POWER/ENTER. The Alarm Time Screen appears.
3 . Use â–²AIR or SHIFT â–¼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
MAN DOWN MENU screen.
Returning to the Maintenance Mode Menu
1 . From the MAN DOWN MENU screen, place the cursor next to
ESCAPE.
2 . Press and release POWER/ENTER. The instrument will return to the
main menu.
Updating Datalogging Parameters
The LOG SETTING menu item in Maintenance Mode has a sub menu with
4 menu items: LOG INTERVAL, LOG CLEAR DISP, LOG OVER
WRITE, and ESCAPE.
1 . From the main menu, place the cursor next to LOG SETTING.
2 . Press and release POWER/ENTER. The LOG SETTING menu
appears.
MAINTE
LOG SETTING
LOG INTERVAL
LOG CLEAR DISP
LOG OVER WRITE
ESCAPE