GX-6000 Operator’s Manual Display Mode 64
1 . With the CAT (LEL) SENSOR PROTECTION screen displayed,
press and release the POWER/ENTER button. The current setting will
be displayed along with a warning about no %LEL alarms if the catalytic
LEL sensor is turned off.
2 . Use the AIR or SHIFT (PANIC) button to display the desired
setting.
NOTE: If the catalytic LEL sensor is disabled and if the instrument is then
turned off and back on, the catalytic LEL sensor will be enabled
again.
3 . Press and release the POWER/ENTER button to return to the CAT
(LEL) SENSOR PROTECTION screen in Display Mode.
LCD Background Color Flip
With LCD BACKGROUND set to ON, the LCD colors will flip. The
background will be black instead of white and the text will be white instead
of black. With LCD BACKGROUND set to OFF (factory setting), the LCD
background will be white and the LCD text will be black.
1 . With the LCD Background Flip Screen showing, press and release the
POWER/ENTER button. The current setting will be displayed.
DISP
CAT (LEL) SENSOR
***WARNING***
NO LEL ALARMS IF
CAT (LEL) IS
DISABLED
ENABLED
DISP
LCD BACKGROUND
YES :ENTER
NO :DISP
EXIT :RESET