GX-6000 Operator’s Manual ESS-03 Calibration • 188
13 . If you are using a 5-gas (standard 4-gas plus ESS-03) calibration
cylinder, continue with Step 14.
If you are using an ESS-03 only calibration cylinder:
a. Connect the tubing from the regulator to the rigid tube on the probe.
Allow the GX-6000 to draw gas for one minute.
b. Press and release the POWER/ENTER button to set the span
adjustment for the ESS-03 channel to the programmed value.
Continue with Step 15.
14 . If you are using a 5-gas (standard 4-gas plus ESS-03) calibration
cylinder:
a. Since calibration gas was flowing to the ESS-03 sensor while the
standard 4 channels were being calibrated, the ESS-03 sensor reading
should be stable and ready for adjustment.
b. Press and release the POWER/ENTER button to set the span
adjustment for the ESS-03 channel to the programmed value.
Continue with Step 15.
15 . If the ESS-03 channel passed calibration, the following screen sequence
occurs.
AUTO
END
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
4:07
HCN
ppm
9.0