EasyManua.ls Logo

RKI Instruments GX-6000 - Page 125

RKI Instruments GX-6000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GX-6000 Operator’s Manual Replacing the Probe’s Particle Filter and Hydrophobic Filter
2 . Grasp each end of the clear probe body firmly and unscrew the two
halves from each other. The top half includes the probe tube and the
bottom half includes the sample hose.
Figure 22: Replacing the Particle Filter and Hydrophobic Filter Disk
3 . Set the top half aside. Make sure the bottom half remains oriented with
the sample hose facing down.
4 . The particle filter should be sitting on top of the hydrophobic filter disk.
If it is not, remove it from the top half of the probe.
5 . Remove the white hydrophobic filter disk from the top of the bottom
half. The disk sits in a black gasket.
6 . Place the new filter disk flat on top of the gasket. Make sure that it sits in
the gasket and does not extend over the gasket’s edge.
7 . Set the new particle filter on top of the filter disk. The bigger end should
be facing down.
Bottom Half
Gasket
Hydrophobic Filter Disk
Sample Hose
Top Half
Particle Filter, Not Used
for Chlorine Detection

Table of Contents

Other manuals for RKI Instruments GX-6000

Related product manuals