EasyManua.ls Logo

RKI Instruments GX-6000 - Page 164

RKI Instruments GX-6000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
155 Using Maintenance Mode GX-6000 Operator’s Manual
3 . Use AIR or SHIFT (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument
will return to the BUMP SETTING screen.
Updating the Bump Expired Setting
This item defines what indication is given during start up when a bump test is
due and BUMP REMINDER is set to ON.
With BUMP EXPRD set to CONFIRM TO USE (factory setting), the GX-
6000 will give an indication at start up if a bump test is past due and will
require the user to decide whether to perform a bump test or continue and use
the GX-6000 without bump testing. Press and release RESET to continue
without calibrating or POWER/ENTER to enter Calibration Mode.
With BUMP EXPRD set to CANNOT USE, if the unit is due for bump
testing, the GX-6000 will give an indication at start up that a bump test is past
due and will prompt you to press and release POWER/ENTER to enter
Calibration Mode and perform a bump test. Using any other button will have
no effect. A successful bump test must be performed in order to use the
instrument.
NOTE: Even if the password function is turned on in PASSWORD, no
password will be required to perform a bump test during startup.
NOTE: If BUMP DISP is set to OFF, the bump test menu item will not
appear in Calibration Mode even though the instrument is
prompting you to perform a bump test. A bump test can always be
performed in Maintenance Mode, if necessary.
With BUMP EXPRD set to NO EFFECT, the GX-6000 will give an
indication at startup that a bump test is past due. If desired, the POWER/
ENTER button can be pressed to enter Calibration Mode and perform a bump
test but it is not necessary to acknowledge the bump test due indication. The
warm-up sequence will continue on its own.
1 . From the BUMP SETTING screen, place the cursor next to BUMP
EXPRD.
2 . Press and release POWER/ENTER. The current Bump Expired setting
will be displayed.
3 . Use AIR or SHIFT (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument
will return to the BUMP SETTING screen.

Table of Contents

Other manuals for RKI Instruments GX-6000

Related product manuals