Roche Diagnostics
144 Instructions for Use · Version 13.0
6 Measurement cobas b 123 POC system
Preanalytics
Dialysis solution
WARNING
Failure to follow sample handling protocols for dialysis solutions may lead to
incorrect patient results
Failure to follow sample handling protocols for dialysis solutions may lead to incorrect
patient results, which may endanger patient lives.
r When collecting dialysis solution to perform a measurement, collect fresh dialysate
with a syringe that does not contain any anticoagulants. Do not collect used dialysate.