cobas b 123 POC system 9 Calibration verification control
Configuring analytical measurement ranges
Configuring analytical measurement ranges
AMRs are ranges of values that an instrument can report directly without needing to
dilute or increase the sample concentration. You may need to create or edit your
instrument’s AMR values based on CVC measurement results or local regulations.
p To add a new AMR
1
Utilities > Configuration > Measurement > Parameter > AMR
2
Press the New button.
Figure 9-5 Configuring AMR screen
3
To select a parameter, press the and buttons on the Parameter panel.
Figure 9-6 Adding an AMR screen