EasyManua.ls Logo

Roche cobas b 123 - Configuring Analytical Measurement Ranges

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
Instructions for Use · Version 13.0 241
cobas b 123 POC system 9 Calibration verification control
Configuring analytical measurement ranges
Configuring analytical measurement ranges
AMRs are ranges of values that an instrument can report directly without needing to
dilute or increase the sample concentration. You may need to create or edit your
instrument’s AMR values based on CVC measurement results or local regulations.
p To add a new AMR
1
Utilities > Configuration > Measurement > Parameter > AMR
2
Press the New button.
Figure 9-5 Configuring AMR screen
3
To select a parameter, press the and buttons on the Parameter panel.
Figure 9-6 Adding an AMR screen

Table of Contents

Related product manuals