EasyManua.ls Logo

Roche cobas b 123 - Sensor Phases

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
42 Instructions for Use · Version 13.0
2 General description cobas b 123 POC system
Consumable information
Sensor phases
After inserting a new Sensor Cartridge, allow up to 60 minutes (typically) for the
cartridge to perform START-UP phase actions.
START-UP phase The START-UP phase begins when a new Sensor Cartridge is inserted into the
system. During the START-UP phase, the system performs actions on the
Sensor Cartridge to prepare it for its first measurement.
These actions are:
o WET-UP procedure: A standby solution moistens the Sensor Cartridge and
activates it at an increased temperature
o Couple check: This ensures that the correct temperatures for BG (37° C) and
ISE/Glu/Lac (30° C) parameters can be maintained
o Swelling time: Time between the end of the couple check and the first system
calibration
o First system calibration: New calibration data are generated during this
calibration. It takes longer than a normal system calibration
RUN-IN phase The RUN-IN phase begins after the START-UP phase and continues for a specific
time, which is defined on the Sensor Cartridge.
During the RUN-IN phase, you can measure without any impact to their accuracy or
precision. However, sample measurement times are longer.
Stable phase The stable phase begins at the end of the RUN-IN phase and ends when the in-use
time of the Sensor Cartridge expires.
IN-USE time The IN-USE time is the time from the beginning of the RUN-IN phase to the end of
the stable phase.
Phase Instrument performance
START-UP phase No measuring operation
RUN-IN phase Measuring operation with longer time-to-display or time-to-
ready
Stable phase Regular operation
IN-USE time Sensor Cartridge in-use time
Table 2-4 Sensor phases

Table of Contents

Related product manuals