EasyManua.ls Logo

Roche cobas b 123 - Performing QC Measurements

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
184 Instructions for Use · Version 13.0
7 Quality control cobas b 123 POC system
Performing QC measurements
Performing QC measurements
You can perform the following types of QC measurements:
o AutoQC measurement
o User-activated AutoQC measurement
o Manual QC measurement
NOTICE
Manual QC measurement
r When performing a manual QC measurement, enter or check the QC ranges from the
package insert.
AutoQC measurements AutoQC measurements are performed automatically at predefined times that you can
configure.
Figure 7-2 QC scheduler
q
The results of an AutoQC measurement are not shown on the screen at the end of a
measurement. You can view the results of AutoQC measurements in the QC database.
u QC database (p. 192)
p To perform a user-activated AutoQC measurement
q
User-activated QC measurements are only possible if the instrument is ready for a
measurement.
1
Workplace > QC measurement
The following window appears on the screen:
Figure 7-3 Quality control menu screen

Table of Contents