EasyManua.ls Logo

Roche cobas b 123 - Performance data

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
Instructions for Use · Version 13.0 55
cobas b 123 POC system 3 Specifications
Performance data
Performance data
Measurement parameters
Precision (sample mode: normal)
“Repeatability (S
R
)” and “Intermediate precision (S
I
)” were determined from 2 runs
per day with 2 replicated per run for 20 days using the cobas b 123 POC system.
The mean value is the measured value of the corresponding parameter for which S
R
and S
I
are representative or have been determined.
Parameter Specified sample types Specified ranges
PO
2
B/Q/C 10–700 mmHg
PCO
2
B/Q/C 10–150 mmHg
pH B/Q/C 6.5–8.0
Na
+
B/Q/C
DS
100–200 mmol/L
120–155 mmol/L
K
+
B/Q/C
DS
1.0–15 mmol/L
1.0–6.0 mmol/L
Cl
-
B/Q/C 70–150 mmol/L
Ca
2+
B/Q/C
DS
0.1–2.5 mmol/L
0.9–1.85 mmol/L
0.4008–10.02 mg/dL
3.6072–7.42 mg/dL
Hct B/Q 10–75%
Glu B/Q/C 1.0–30 mmol/L 18.016–540.48 mg/dL
Lac B/Q/C 1.0–20 mmol/L 9.0080–180.16 mg/dL
tHb (COOX) B/Q 4–25 g/dL
SO
2
(COOX) B/Q 30–100%
HHb (COOX) B/Q 0–70%
COHb (COOX) B/Q 0–70%
O
2
Hb (COOX) B/Q 30–100%
MetHb (COOX) B/Q 0–70%
Bili (neonatal) (COOX) B/Q 3–50 mg/dL 51.3–855 mol/L
Baro 530–800 mmHg
Table 3-1 Measurement parameters
B
Whole blood
DS
Dialysis solution
Q
Aqueous QC material
(1)
(1) with approximate physiological ion matrix and buffer capacity.
C
Aqueous solution

Table of Contents