Instructions for Use · Version 13.0 321
cobas b 123 POC system 12 Troubleshooting
War nings
Warnings
Warnings are advance notifications of possible issues with the instrument or
consumables. You can perform measurements, but you may not be able to perform
specific functions if it involves the affected module of the instrument or the
consumable.
To remedy a warning, follow the instructions on the screen.
q
If you have to resolve a warning and it is not listed, contact your Roche Service
representative.
ID Category Cause Action
5000 Fluid Pack warning Critical fill level in Fluid Pack o Press the Change Fluid Pack button. Follow the
instructions on the screen.
o If this error persists, contact your Roche Service
representative.
5001 Fluid Pack warning Fluid Pack nearly expired
o Press the Change Fluid Pack button. Follow the
instructions on the screen.
o If this error persists, contact your Roche Service
representative.
5002 AutoQC Pack warning Critical fill level in AutoQC Pack
o Press the Change AutoQC Pack button. Follow the
instructions on the screen.
o If this error persists, contact your Roche Service
representative.
5003 AutoQC Pack warning AutoQC Pack nearly expired
o Press the Change AutoQC Pack button. Follow the
instructions on the screen.
o If this error persists, contact your Roche Service
representative.
5004 Sensor Cartridge warning Total number of tests close to limit
o Press the Change Sensor Cartridge button. Follow
the instructions on the screen.
o If this error persists, contact your Roche Service
representative.
5005 Sensor Cartridge warning Sensor Cartridge nearly expired
o Press the Change Sensor Cartridge button. Follow
the instructions on the screen.
o If this error persists, contact your Roche Service
representative.
5006 Temperature warning Temperature range exceeded
o Press the Restart system button.
o If this error persists, contact your Roche Service
representative.
5007 Instrument warning Magnetic switch defective
o Contact your Roche Service representative.
Table 12-3 Warnings