Instructions for Use · Version 13.0 85
cobas b 123 POC system 3 Specifications
Environmental parameters
Fluid Pack
QC material
Operating conditions
o Ambient temperature up to 42 days at 15-32 °C
o Relative humidity 15–90% (not condensed)
Table 3-122
Storage conditions in original packaging
o Temperature up to 9 months at 15-25 °C
o Relative humidity 15–85% (not condensed)
Table 3-123
Stability during operation
o cobas b 123 Fluid Pack COOX 200
up to 200 tests or up to 42 days
(1)
(1) All Fluid Packs can be used for up to 42 days. The indicated number of tests will be achieved when a
minimum number of measurements are performed per day.
o cobas b 123 Fluid Pack COOX 400
up to 400 tests or up to 42 days
(1)
o cobas b 123 Fluid Pack COOX 700
up to 700 tests or up to 42 days
(1)
o cobas b 123 Fluid Pack 200
up to 200 tests or up to 42 days
(1)
o cobas b 123 Fluid Pack 400
up to 400 tests or up to 42 days
(1)
o cobas b 123 Fluid Pack 700
up to 700 tests or up to 42 days
(1)
Table 3-124
Storage conditions in original packaging
Temperature
o COMBITROL PLUS B (Level 1–3)
o cobas b 123 AutoQC Pack
o cobas b 123 AutoCVC Pack
o up to 24 months at 2-8 °C
o up to 24 months at 2-8 °C
o up to 24 months at 2-8 °C
Relative humidity 15–85% (not condensed)
Table 3-125
Stability during operation
o COMBITROL PLUS B (Level 1–3) up to 3 months at room temperature (up to 28 °C)
o cobas b 123 AutoQC Pack up to 3 months in the instrument
o cobas b 123 AutoCVC Pack up to 3 months in the instrument
Table 3-126