EasyManua.ls Logo

Roche cobas b 123 - Fluid Pack

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
46 Instructions for Use · Version 13.0
2 General description cobas b 123 POC system
Consumable information
Fluid Pack
Figure 2-23 cobas b 123 Fluid Pack
The Fluid Pack contains:
o 7 bags:
O 2 waste water containers
O 1 reference solution
O 1 wetting solution (to wet the Sensor Cartridge)
O 1 standby solution (STDBY) (for recalibration and standby calibration)
O Calibration solution CAL 1
O Calibration solution CAL 2
o Sample input module (fill port and needle)
o Tubi ng
o Cuvette for oximeter module (optional)
An air filter is integrated in the base of the Fluid Pack to control the temperature of
the system’s modules.
Fluid Pack types Different Fluid Packs are available to support your testing frequency.
A Tubing (peristaltic pump) D Docking parts (measuring chamber module)
B Cuvette (oximeter module) E Sample input module
C Sample sensor contacts
A
D
B
C
E

Table of Contents

Related product manuals