EasyManuals Logo

Roche cobas b 123 User Manual

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #230 background imageLoading...
Page #230 background image
Roche Diagnostics
230 Instructions for Use · Version 13.0
8 Calibration cobas b 123 POC system
Configuring calibration settings
Configuring calibration settings
You can configure when your instrument performs 2P and system calibrations in the
Calibration scheduler. In addition, you can activate or deactivate automatic printing
of calibration reports.
p To define the system calibration’s start time
1
Utilities > Configuration > Calibration > Times / intervals
2
To define the system calibration’s start time, press the button on the
System calibration panel.
Figure 8-4 Calibration scheduler screen
The screen displays a numeric scroll box.
Figure 8-5 Numeric scroll box for defining the system calibration’s start time
3
To set the correct start time, press the and buttons on the Start time panel.
Then, press the OK button.
s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 123 and is the answer not in the manual?

Roche cobas b 123 Specifications

General IconGeneral
BrandRoche
Modelcobas b 123
CategoryMedical Equipment
LanguageEnglish

Related product manuals