cobas b 123 POC system 11 Consumable change
Disinfecting the instrument
Disinfecting the instrument
Regularly disinfect the following parts of the instrument:
o Touch screen
o Surfaces of the instrument (including all covers)
o Consumable area
p To disinfect the instrument’s screen
NOTICE
Spraying water or disinfectant onto the screen may damage the instrument
Disinfecting the instrument’s screen by directly spraying water, disinfectants, or other
liquids may damage the instrument.
r When using liquids to disinfect the screen, first spray the liquid onto a damp cloth.
Then, wipe the desired location with this cloth.
r Do not directly spray any liquids onto the screen.
1
Instrument > Disinfection of touch screen
The touch screen is deactivated for 30 seconds.
2
Clean the screen with a damp cloth moistened with disinfectant.
s
p To disinfect the instrument’s surfaces
NOTICE
Spraying disinfectant onto the instrument may damage the instrument
Spraying disinfectant onto the instrument may damage the instrument’s electronics.
r When using disinfectants, first spray the disinfectant onto a damp cloth. Then, wipe the
desired location with this cloth.
r Do not spray disinfectant on parts that cannot be removed, or that are inside the
instrument.
1
Clean all surfaces with a damp cloth moistened with disinfectant.
If there are locations that need more cleaning, clean these surfaces first with a
swab or cellulose sponge moistened with distilled water, and then clean with a
damp cloth moistened with disinfectant.
WARNING
Immediate use after disinfection may lead to fire or explosion
Traces of disinfectant on the instrument after a disinfection may provoke a fire or
explosion.
r After disinfecting the instrument, allow the disinfectant to evaporate from the
instrument for 15 minutes.
2
Allow the disinfectant to evaporate from the surfaces for 15 minutes.