Roche Diagnostics
302 Instructions for Use · Version 13.0
11 Consumable change cobas b 123 POC system
Disinfecting the instrument
p To disinfect the instrument’s consumable area
WARNING
Disinfecting the Sensor Cartridge or the measuring chamber module may lead to
incorrect patient results
Disinfecting the Sensor Cartridge or the measuring chamber module may deteriorate the
performance of the Sensor Cartridge’s sensors, which may lead to incorrect patient results.
r Do not disinfect the Sensor Cartridge or the measuring chamber module. Do not
disinfect any part that cannot be removed, or that are inside the instrument.
r Do not spray any water or disinfectants on the Sensor Cartridge, or inside the
measuring chamber module.
CAUTION
Displaced valve levers
Displaced valve levers cause problems when inserting the Fluid Pack.
r During disinfection, be absolutely certain not to displace the valve levers.
NOTICE
Disinfecting the consumables area with water or sprays may damage the
instrument’s electronics
Disinfecting the consumables area by directly spraying water, disinfectants, or other liquids
may damage the instrument’s electronics.
r When using disinfectants, first spray the disinfectant onto a damp cloth. Then, wipe the
desired location with this cloth.
r Do not spray disinfectant on parts that cannot be removed, or that are inside the
instrument.
q
Disinfect the instrument’s consumable area as part of each consumable change and the
put out of operation routine. Do not disinfect the instrument’s consumable area at any
other time.
1
Clean the consumable area with a damp cloth moistened with disinfectant.
To protect the Sensor Cartridge and the hemolyzer, avoid disinfecting the left wall
of the consumable area. The left wall is the separation between the measuring
chamber and oximeter modules.
If there are locations that need more cleaning, clean these surfaces first with a
swab or cellulose sponge moistened with distilled water, and then clean with a
damp cloth moistened with disinfectant.
WARNING
Immediate use after disinfection may lead to fire or explosion
Traces of disinfectant on the instrument after a disinfection may provoke a fire or
explosion.
r After disinfecting the instrument, allow the disinfectant to evaporate from the
instrument for 15 minutes.
2
Allow the disinfectant to evaporate from the surfaces for 15 minutes.
s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 123 and is the answer not in the manual?

Roche cobas b 123 Specifications

General IconGeneral
BrandRoche
Modelcobas b 123
CategoryMedical Equipment
LanguageEnglish

Related product manuals