Roche Diagnostics
228 Instructions for Use · Version 13.0
8 Calibration cobas b 123 POC system
User-activated calibrations
User-activated calibrations
User-activated calibrations are available in the Instrument tab.
Figure 8-2 Instrument tab
You can perform the following calibrations manually:
o Calibration for “Ready”
o 1-point calibration
o 2-point calibration
o System calibration
o Oximeter calibration
(1)
u Automatic calibrations (p. 225)
q
Manual calibrations can postpone scheduled automatic calibrations.
You can stop a 2P calibration or system calibration by pressing the button. 1P
calibrations cannot be stopped by pressing the button.
Calibration for “Ready”
The system automatically selects a calibration that sets all measurement parameters
to the “Ready” state.
(1) For instruments with COOX module only.