Roche cobas b 123 - Comparison of Micro Samples to Normal Samples; Glu; Lac

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
78 Instructions for Use · Version 13.0
3 Specifications cobas b 123 POC system
Performance data
Glu
Unit: [mmol/L]
Material: whole blood
Lac
Unit: [mmol/L]
Material: whole blood
Comparison of micro samples to normal samples
The trueness of measurement values in micro sample mode, BG-only (40 L) and BG
with COOX (55 L), have been confirmed by comparing measurement values in
normal sample mode (150 L).
Number of instruments: 8 cobas b 123 POC systems
Material: whole blood
Comparison instrument No. of samples [n] Bias
(1)
Slope [b] Intercept [a] Pearson’s Corr. coeff. [r]
cobas® 6000 686 -2.4 0.98 -0.04 0.98
Table 3-110 Glu
(1) Median absolute bias
Comparison instrument No. of samples [n] Bias
(1)
Slope [b] Intercept [a] Pearson’s Corr. coeff. [r]
cobas® 6000 590 --- 1.00 0.06 0.99
Table 3-111 Lac
(1) Median absolute bias
Parameter Unit No. of samples [n] Bias
(1)
Slope [b] Intercept [a] Pearson’s Corr. coeff. [r]
pH [pH units] 149 0.00 0.9932 0.0506 0.9996
PO
2
[mmHg] 205 0.564 0.994 1.0251 0.9995
PCO
2
[mmHg] 133 0.09 0.9646 1.4684 0.9989
Hct [%] 370 -0.026 1.0034 -0.1195 0.9988
tHb [g/dL] 244 -0.044 1.033 -0.4168 0.9998
O
2
Hb
[%] 403 0.571 0.9806 2.1444 0.9993
HHb [%] 403 -0.558 0.9799 -0.2087 0.9993
MetHb [%] 176 -0.332 0.9948 -0.0827 0.9999
COHb [%] 278 -0.034 1.0001 -0.038 1.0000
SO
2
[%] 484 0.431 0.9861 1.572 0.9992
Bili [mg/dL] 232 -0.506 0.9975 -0.4805 0.9982
Table 3-112 Sample comparison data
(1) Median absolute bias