cobas b 123 POC system 1 Safety information
Safety precautions
Infectious waste Contact with waste (liquid and/or solid) may result in infection. All materials and
mechanical components associated with the waste systems are potentially
biohazardous.
o Wear appropriate personal protective equipment. Take extra care when working
with lab gloves. They can be easily pierced or cut, leading to infection.
o If any biohazardous material is spilled, wipe it up immediately and apply a
disinfectant.
o If waste comes into contact with your skin, wash the affected area immediately
with soap and water and apply a disinfectant. Consult a physician.
Environmental harm The system generates liquid and/or solid waste. This waste contains concentrated
reaction solutions and is potentially biohazardous. Improper disposal may
contaminate the environment.
o Treat this waste as infectious waste.
o Dispose of waste in accordance with the local regulations.
Mechanical safety
CAUTION
Opening and closing the instrument’s door
Opening and closing the instrument’s door may result in personal injury to your hands.
r Always open and close the door with both hands.
r When opening and closing the door, do not hold other objects.
Consumables
WARNING
Incorrect handling of consumables
Incorrect handling of the consumables may result in infection, and errors in diagnosis, both
posing danger to the patient.
r Always handle consumables correctly. The Fluid Pack has to be carried by its handle in
an upright position.
r Do not use consumables that have exceeded their expiry dates. The expiry date of
consumables is shown on its packaging.
r When handling used consumables, observe relevant laboratory procedures on safe
usage. Wear appropriate personal protective equipment, such as:
o Lab gloves
o Lab coat
o Eye protection
o Face shield if there is a chance of splashing or splattering
r Dispose of consumables according to the local regulations. For more information,
contact your Roche service representative
WARNING
Handling QC material and AutoCVC packs
r QC material and AutoCVC packs must have adjusted to room temperature for at least
24 hours before performing a measurement.