EasyManuals Logo

Roche cobas b 123 User Manual

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #183 background imageLoading...
Page #183 background image
Roche Diagnostics
Instructions for Use · Version 13.0 183
cobas b 123 POC system 7 Quality control
Important information for evaluating QC results
Important information for evaluating QC results
QC measurement results are evaluated based on their ranges:
o Measured value is within the target value range ± 2
O The QC result is acceptable and the parameter is/remains active for
measurements.
o Measured value is outside the target value range ± 2
O Consequence: A QC consequence is assigned to the parameter.
O The QC result is not acceptable. The parameter is locked for further
measurements. It is unlocked for further measurements when the cause of the
QC consequence has been determined and the error has been resolved.
O Repeat the QC measurement.
u Removing QC consequences (p. 215)