12 Troubleshooting cobas b 123 POC system
Information alarms
Information alarms
Information alarms refer to actions that you have to perform. You can perform all
functions with the instrument while information alarms are active.
To remedy an information alarm, follow the instructions on the screen.
q
If you have to resolve an information alarm and it is not listed, contact your Roche Service
representative.
ID Category Cause Action
7000 Service action Deleting of instrument database
requested
o To remove all data and configurations from the
instrument database, press the Restart system
button.
o To abort the request, press the Cancel button.
7001 Service action Software update pending
o To install the software update, press the
Restart system button.
o To abort the software update, press the Cancel
button.
7002 Service action Configuration file pending
o To apply the new configuration, press the
Restart system button.
o To abort the request, press the Cancel button.
Table 12-4 Information alarm