EasyManua.ls Logo

Roche cobas b 123 - Bili; Precision (Sample Mode: Micro Sample)

Roche cobas b 123
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
66 Instructions for Use · Version 13.0
3 Specifications cobas b 123 POC system
Performance data
Material: cobas b 123 AutoQC Pack TRI-LEVEL, n=80
Material: cobas b 123 AutoCVC Pack, n=80
Bili
Unit: [mg/dL]
Material: cobas b 123 AutoQC Pack TRI-LEVEL, n=80
Material: cobas b 123 AutoCVC Pack, n=80
Material: prepared whole blood, including bilirubin, n=40
Precision (sample mode: micro sample)
“Repeatability (S
R
)” and “Intermediate precision (S
I
)” were determined using the
cobas b 123 POC system for 10 days.
The mean value is the measured value of the corresponding parameter for which S
R
and S
I
are representative or have been determined.
Tonometered whole blood Whole blood was tonometered at 37°C with high-precision gas mixtures.
Mean S
R
(CV%) S
I
(CV%)
Level 1 17.6 0.0414 0.24 0.0909 0.52
Level 2 8.2 0.1451 1.78 0.1651 2.02
Level 3 2.7 0.1279 4.73 0.1329 4.92
Table 3-58 HHb material
Mean S
R
(CV%) S
I
(CV%)
Level 5 19.7 0.0410 0.21 0.0723 0.37
Level 6 1.9 0.1319 6.88 0.1260 6.58
Table 3-59 HHb material
Mean S
R
(CV%) S
I
(CV%)
Level 1 5.8 0.0247 0.43 0.1176 2.03
Level 2 11.7 0.0988 0.84 0.1532 1.31
Level 3 20.2 0.2123 1.05 0.2663 1.32
Table 3-60 Bili material
Mean S
R
(CV%) S
I
(CV%)
Level 5 4.7 0.0183 0.39 0.0915 1.96
Level 6 22.2 0.1650 0.74 0.2292 1.03
Table 3-61 Bili material
Mean S
R
(CV%) S
I
(CV%)
Sample 1 5.3 0.113 2.12 0.245 4.60
Sample 2 14.74 0.132 0.90 0.255 1.73
Sample 3 23.20 0.111 0.48 0.389 1.68
Table 3-62 Bili material

Table of Contents