Roche Diagnostics
134 Instructions for Use · Version 13.0
5 Installation and put out of operation cobas b 123 POC system
Put out of operation
Put out of operation
For less than 24 hours
If the instrument is not needed for less than 24 hours, switch it off by pressing the
following buttons: Utilities > Switch off. Follow the instructions on the screen.
WARNING
Instrument is off for more than 1 hour
r Perform a system calibration and QC measurements on 3 levels when you turn the
instrument on again.
If the instrument is off for less than 1 hour, perform a 2P calibration when you turn
the instrument on again.
q
Do not use previously installed Fluid Packs and Sensor Cartridges that have been outside
of an instrument for more than 24 hours.
Do not use previously installed AutoQC Packs that have been outside of an instrument for
more than 7 days.
Do not use previously installed AutoQC Packs, Fluid Packs and Sensor Cartridges that
have exceeded the following cumulative offboard times:
o AutoQC Packs: 7 days
o Fluid Packs: 40 days
o Sensor Cartridges: 10 days
q
To disconnect the instrument from the power supply, the power supply has to be
disconnected from the power socket.
For more than 24 hours
If the instrument is not needed for more than 24 hours, put the instrument out of
operation.